Amperometric immunosensor for nonylphenol determination based on peroxidase indicating reaction

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
9
0

Year Published

2009
2009
2017
2017

Publication Types

Select...
6
3

Relationship

0
9

Authors

Journals

citations
Cited by 34 publications
(9 citation statements)
references
References 25 publications
0
9
0
Order By: Relevance
“…A limit of detection as low as 1.0×10 −8 mol·L −1 is estimated (S/N = 3). Compared with other electrochemical sensors, such as 3.2 × 10 −7 mol·L −1 for TiO 2 -nano Au-molecularly imprinted sensor [19], 2.6 × 10 −7 mol·L −1 for NiTPPS/CNTs sensor [27] and 10 μg·L −1 (≈4.5 × 10 −8 mol·L −1 ) for HRP-based immunosensor [28], this is a competitive value.…”
Section: Resultsmentioning
confidence: 92%
“…A limit of detection as low as 1.0×10 −8 mol·L −1 is estimated (S/N = 3). Compared with other electrochemical sensors, such as 3.2 × 10 −7 mol·L −1 for TiO 2 -nano Au-molecularly imprinted sensor [19], 2.6 × 10 −7 mol·L −1 for NiTPPS/CNTs sensor [27] and 10 μg·L −1 (≈4.5 × 10 −8 mol·L −1 ) for HRP-based immunosensor [28], this is a competitive value.…”
Section: Resultsmentioning
confidence: 92%
“…The linear ranges and detection limits obtained for TOP, OP, and NP determination by DPV on PGE are similar to or even better than (e.g., TOP) some others presented in the literature. 10,12,16,17 The described PGE has the main advantage of a cheap, simple, and fast electrode surface "regeneration", which enhances its applicability to routine analysis and offers the possibility of direct on-field testing by using a portable potentiostat. The PGE constitutes, thus, a less expensive alternative to commercially available disposable screen-printed electrodes.…”
Section: Discussionmentioning
confidence: 99%
“…These results are similar to or even better than (e.g., for TOP) those reported in the literature for the electrochemical detection of different APs using various working electrodes. 10,12,16,17 Selectivity: Common metallic ions like Cu 2+ , Pb 2+ , and Cd 2+ , which can exist in water samples, do not interfere in the voltammetric determination of APs because as cations they are not oxidized voltammetrically. DPV studies performed in PBS (pH 7.40) at activated PGE on a mixture of common PAHs did not show any voltammetric signal in the potential range of 0.0 to 1000 mV, such that neither of these compounds interfere in AP voltammetric determination on PGE.…”
Section: Validation Of the Methodsmentioning
confidence: 99%
“…Generally, alkylphenols are detected by liquid chromatography [5][6][7][8] and gas chromatography. [9][10][11] Although electrochemical methods were also exploited for the determination of alkylphenols, most of them aimed at nonylphenol [12][13][14] and just a few concerned other alkylphenols. As examples, Ana-Maria Gurban et al 15 developed an alkylphenol biosensor by entrapping horseradish peroxidase in a nanocomposite gel of single-walled carbon nanotubes and 1-butyl-3-methylimidazolium hexauorophosphate ionic liquid.…”
Section: Introductionmentioning
confidence: 99%