Abstract:We designed and implemented a two-axis measurement system to inspect the assembly quality condition of a bone conduction implant (BCI) transducer. The system consists of a laser Doppler vibrometer (LDV), XY manual stage, and two digital scale calipers capable of displaying a position coordinate system. To measure the vibration of the cantilever constituting the transducer vibrational membrane, an XY coordinate system was obtained using FEA software. Based on the derived XY coordinate system, the cantilever vib… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.