An in vitro study on the impact of the torque after repeated use of the implant handpiece: An original research

Abstract: Introduction: Extreme repeated use of the dental implant handpiece may cause aging of the handpiece through heat generation of the implant motor and affect its lifespan. We aim to conduct an in-vitro study on the impact of the torque after repeated use of the implant handpiece.  Materials and methods: Two types of implant handpiece systems NSK, and SAESHIN were applied. The output torque was calibrated using a digital torque gauge. The experiment was done under the setting torque value of 35 Ncm (implant place… Show more