2018
DOI: 10.11144/javeriana.sc23-1.aint
|View full text |Cite
|
Sign up to set email alerts
|

An investigation on the effect of humidity on the zero signal of a strain gauge measuring system

Abstract: Humidity is one of the most important ambient factors influencing torque calibration readings. The present study addresses the effect of relative humidity on the readings of reference torque transduction and amplification devices. The results revealed a linear effect of increasing humidity on the readings of a DMP40 torque amplifier and a torque transducer at four different temperatures. Furthermore, humidity effects were smaller on the DMP40 amplifier than on the torque transducer across all four temperatures… Show more