Analysis and quantification of vitamin D metabolites in serum by ultra‐performance liquid chromatography coupled to tandem mass spectrometry and high‐resolution mass spectrometry – a method comparison and validation

Abstract: Quantification of the relevant vitamin D metabolites was successfully developed and validated here. It was shown that LC/HRMS is an accurate, powerful and easy to use approach for quantification within clinical laboratories. Finally, the results here suggest that it is important to separate 3-epi-25OH-D3 from 25OH-D3.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

3
44
1

Year Published

2013
2013
2022
2022

Publication Types

Select...
7
1

Relationship

1
7

Authors

Journals

citations
Cited by 74 publications
(48 citation statements)
references
References 38 publications
3
44
1
Order By: Relevance
“…Nevertheless, Bland-Altman analysis did not show any concentration-dependent or systematic differences between HRMS and DiaSorin (Figure 3d). These results are mostly in agreement with findings from other studies using Orbitraps for HRMS, where generally strong correlations between Orbitrap and QqQ [15], and Orbitrap and IDS-iSYS 25-hydroxyvitamin D immunoassay [18] were shown.…”
Section: Clinical Samples From Patient Studysupporting
confidence: 92%
See 2 more Smart Citations
“…Nevertheless, Bland-Altman analysis did not show any concentration-dependent or systematic differences between HRMS and DiaSorin (Figure 3d). These results are mostly in agreement with findings from other studies using Orbitraps for HRMS, where generally strong correlations between Orbitrap and QqQ [15], and Orbitrap and IDS-iSYS 25-hydroxyvitamin D immunoassay [18] were shown.…”
Section: Clinical Samples From Patient Studysupporting
confidence: 92%
“…Although narrowing the window size clearly reduced the abundances of isobars in the TOF traces, the problem of isobaric cross-talk persisted, with resulting unsatisfactory signal-to-noise ratios and insufficient precision. Clearly, the higher resolving power obtained from an Orbitrap would be beneficial in this situation [15,18]. There was little influence on peak areas in the CID-MS/MS traces when the extraction size was reduced, however, as isobaric noise generally was very low in the CID spectra (Figure 1d).…”
Section: Hrms: Tof-ms Versus Tof-ms/msmentioning
confidence: 99%
See 1 more Smart Citation
“…A fast, accurate, and reliable method to quantify the vitamin D metabolites-25-hydroxyvitamin D 2 , 25-hydroxyvitamin D 3 , and 3-epi-25hydroxyvitamin D 3 -in human serum samples was developed and validated for this project (21). An ultrahigh-pressure liquid chromatography-tandem mass spectrometry system was developed.…”
Section: (Oh)dmentioning
confidence: 99%
“…Previously, the quantitative aspects of low resolution MS (LR-MS) operated in SRM mode has been compared with several different modes in HR-MS [21][22][23][24][25][26][27][28]. Comparisons of selectivity were less often subject of investigation.…”
Section: Introductionmentioning
confidence: 99%