2018
DOI: 10.1016/j.endinu.2017.11.003
|View full text |Cite
|
Sign up to set email alerts
|

Analytical interference in the corticotropin immunoassay in patients with adrenal adenomas

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4

Citation Types

0
5
0

Year Published

2019
2019
2022
2022

Publication Types

Select...
3

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(5 citation statements)
references
References 10 publications
0
5
0
Order By: Relevance
“…In all but one of the ACTH assay interference cases previously reported, the Siemens Immulite platforms were used [3][4][5][7][8][9][10] (Table 3), whereas in the present case, the Roche Diagnostics Elecsys ACTH assay yielded falsely elevated ACTH levels, for the first time to our knowledge. The Siemens Immulite platforms adopt a murine monoclonal anti-ACTH antibody and a rabbit polyclonal antibody that is considered capable of react-ing with unknown interference antibodies [9]. On the other hand, the Elecsys ACTH assay is an electrochemiluminescence method employing two monoclonal antibodies specific for ACTH (9-12) and for ACTH (36-39).…”
Section: Discussionmentioning
confidence: 57%
See 3 more Smart Citations
“…In all but one of the ACTH assay interference cases previously reported, the Siemens Immulite platforms were used [3][4][5][7][8][9][10] (Table 3), whereas in the present case, the Roche Diagnostics Elecsys ACTH assay yielded falsely elevated ACTH levels, for the first time to our knowledge. The Siemens Immulite platforms adopt a murine monoclonal anti-ACTH antibody and a rabbit polyclonal antibody that is considered capable of react-ing with unknown interference antibodies [9]. On the other hand, the Elecsys ACTH assay is an electrochemiluminescence method employing two monoclonal antibodies specific for ACTH (9-12) and for ACTH (36-39).…”
Section: Discussionmentioning
confidence: 57%
“…It is extremely rare that falsely high or low ACTH levels by a substance in the specimen interfering with the ACTH measurements are confirmed. There have been only 23 such cases, to our knowledge [3][4][5][6][7][8][9][10] (Table 3). Of the 23 cases, 22 exhibited erroneously elevated ACTH levels.…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…Previous studies have illustrated that the interaction between these antibodies and other unknown interference antibodies results in most of the doubtful plasma ACTH elevations analyzed using the ACTH (Immulite). 12 Cisbio ELSA-ACTH assay used at the LEZEN Reference Lab was composed of an ACTH N-terminal and C-terminal specific monoclonal antibody, which were, respectively, coated on the ELSA solid phase and radiolabeled with iodine-125 tracer. The increased specificity of the ELSA-ACTH immunoassay yields more accurate plasma ACTH levels.…”
Section: Discussionmentioning