Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter

Abstract: Purpose: The purpose of this study is to automate the slice thickness verification on the AAPM CT performance phantom and validate it for variations of slice thickness, position from iso-center, and reconstruction filter.Methods: An automatic procedure for slice thickness verification on AAPM CT performance phantom was developed using MATLAB R2015b. The stair object image within the phantom was segmented, and the middle stair object was located. Its angle was determined using the Hough transformation, and the … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
18
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
4
2

Relationship

3
3

Authors

Journals

citations
Cited by 15 publications
(18 citation statements)
references
References 26 publications
0
18
0
Order By: Relevance
“…Most often, a thicker slice increases the contrast of the image and might provide better identification of the volume of interest 12 . However, previous studies have investigated the optimal slice thickness for different purposes and the results have not been consistent 11 – 16 . These studies used the same platform as in the present study 18 – 20 ; however, only one mentioned a slice thickness of 1 mm for CBCT measurements 18 .…”
Section: Discussionmentioning
confidence: 97%
See 3 more Smart Citations
“…Most often, a thicker slice increases the contrast of the image and might provide better identification of the volume of interest 12 . However, previous studies have investigated the optimal slice thickness for different purposes and the results have not been consistent 11 – 16 . These studies used the same platform as in the present study 18 – 20 ; however, only one mentioned a slice thickness of 1 mm for CBCT measurements 18 .…”
Section: Discussionmentioning
confidence: 97%
“…CBCT scans depend on many parameters, such as voxel size, field of view, electric current intensity, reconstruction algorithm, beam collimation and slide thickness 11 . Slice thickness affects the resolution of the image, leading to an impact on the accuracy of measurements 11 . The slice thickness is inversely proportional to image noise and decreasing the slice thickness increases the latter 14 .…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…13 Sofiyatun et al 13 proposed an algorithm to measure slice thickness using the AAPM CT performance phantom automatically. The method was then evaluated for variations of reconstruction filter and distance from the iso-center by Lasiyah et al 14 Results were less than 1.0 mm different from the nominal slice thickness under a variety of conditions (i.e. slice thickness, distance to iso-center, angle of rotation, and filter).…”
Section: Introductionmentioning
confidence: 99%