Characterization of a commercial plastic scintillator for electron FLASH dosimetry
Kyuhak Oh,
Megan A. Hyun,
Kyle J. Gallagher
et al.
Abstract:PurposeThis study investigated the potential of a commercially available plastic scintillator, the Exradin W2, as a real‐time dosimeter for ultra‐high‐dose‐rate (UHDR) electron beams. This work aimed to characterize this system's performance under UHDR conditions and addressed limitations inherent to other conventional dosimetry systems.Methods and materialsWe assessed the W2's performance as a UHDR electron dosimeter using a 16 MeV UHDR electron beam from the FLASH research extension (FLEX) system. Additional… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.