Comparative Effectiveness of Two Models of Point-of-Care Ultrasound for Detection of Post-Void Residual Urine during Acute Ischemic Stroke: Preliminary Findings of Real-World Clinical Application
Abstract:We conducted a comparative study of two models of point-of-care ultrasound devices for measuring post-void residual urine (PVRU). We prospectively enrolled 55 stroke inpatients who underwent both real-time B-mode ultrasound (Device A) and automated three-dimensional (3D) scanning ultrasound (Device B), with a total of 108 measurements. The median PVRU volume of Device B was 40 mL larger than that of Device A. The PVRU difference between the devices was positively and linearly correlated with PVRU. The correlat… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.