Comparison of model-based iterative vs non-iterative 3D device reconstruction for continuous-sweep limited-angle fluoroscopy
Martin G. Wagner,
Ayca Z. Kutlu,
Brian Davis
et al.
Abstract:Continuous-sweep limited angle (CLA) fluoroscopy is a recently proposed method for live 3D catheter reconstruction using a single-plane C-arm that continuously rotates back-and-forth within a narrow angle during fluoroscopic image acquisition. This study compares the accuracy and computation time of two real-time 3D catheter reconstruction algorithms. The first approach, iterative model-based device reconstruction (MDR), optimizes the 3D location of 8 control points representing the catheter as a cubic spline.… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.