2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC) 2012
DOI: 10.1109/nssmic.2012.6551880
|View full text |Cite
|
Sign up to set email alerts
|

Complete scheme for beam hardening correction in small animal computed tomography

Abstract: Abstract-. CT images are often affected by beam hardening artifacts due to the polyenergetic nature of the X-ray beam.Several correction methods have been proposed to be included in an FDK reconstruction scheme, which is still the prefered reconstruction algorithm in commercial X-ray CT scanners.In this work, we present a complete correction scheme for beam hardening artifact correction in an FDK-based reconstruction scenario that accounts for both cupping artifact and dark streaks. The proposed method substit… Show more