Abstract: RÉSUMÉABSTRACT The metrological parameters of a ferrous sulphate dosemeter are normally determined by comparison with primary standards. We have undertaken to evaluate these parameters using a secondary standard and the IAEA formalism for absorbed dose determination. Thus, we have determined with measurements the wavelength of absorption of the ferric ion (304 nm), the molar extinction coefficient e (218.7 ± 0.9) mol 1 m 2 and the radiochemical yield G (1.627 ± 0.04) 1(H> mol J" 1 . These values of c and G are… Show more