2009
DOI: 10.3348/kjr.2009.10.3.217
|View full text |Cite
|
Sign up to set email alerts
|

Coronary Artery Stent Evaluation Using a Vascular Model at 64-Detector Row CT: Comparison between Prospective and Retrospective ECG-Gated Axial Scans

Abstract: ObjectiveWe wanted to evaluate the performance of prospective electrocardiogram (ECG)-gated axial scans for assessing coronary stents as compared with retrospective ECG-gated helical scans.Materials and MethodsAs for a vascular model of the coronary artery, a tube of approximately 2.5-mm inner diameter was adopted and as for stents, three (Bx-Velocity, Express2, and Micro Driver) different kinds of stents were inserted into the tube. Both patent and stenotic models of coronary artery were made by instillating … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

1
4
0

Year Published

2010
2010
2021
2021

Publication Types

Select...
3
1

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(5 citation statements)
references
References 18 publications
1
4
0
Order By: Relevance
“…Consistent with previous studies ( 14 , 15 , 20 ), our study demonstrated that the measurements of coronary artery stent images from PTA acquisition is significantly better than that from RGH. There are two potential explanations of the artifact reductions by PTA scans.…”
Section: Discussionsupporting
confidence: 92%
See 1 more Smart Citation
“…Consistent with previous studies ( 14 , 15 , 20 ), our study demonstrated that the measurements of coronary artery stent images from PTA acquisition is significantly better than that from RGH. There are two potential explanations of the artifact reductions by PTA scans.…”
Section: Discussionsupporting
confidence: 92%
“…The artificial lumen narrowing (ALN) was calculated with the following equation: ALN = (ISD - ISD measured )/ISD ( 14 , 15 ). The artificial attenuation increase (AAI) between in-stent and in-vessel were also calculated with the following equation: AAI = AV in-stent - AV in-vessel .…”
Section: Methodsmentioning
confidence: 99%
“…All measurements were repeated three times and the mean of the measurements by two observers was calculated separately. Artificial lumen narrowing was calculated using the following equation [9,10]: artificial lumen narrowing=(inner stent diameter -visible lumen diameter )/inner stent diameter.…”
Section: Measurement Of Artificial Lumen Narrowingmentioning
confidence: 99%
“…Although retrospective electrocardiographically (ECG)-gated helical (REH) scans are mainly used, with advances in technique, prospective ECG-triggered axial (PEA) scans allow a substantial reduction in radiation dose compared with REH scans (15 mSv reduced to 1.0-5 mSv) while having a possible advantage for in-stent restenosis imaging owing to the reduction of artefacts compared with REH scans [6][7][8][9][10]. Therefore, the purpose of our study was to clarify the influence of scan modes and tube voltage on the in-stent visibility, artificial lumen narrowing and lumen attenuation by using in-stent 50% and 75% restenosis vessel models.…”
Section: Introductionmentioning
confidence: 99%
“…In the CCTA image, the metal artifact of the scaffold has the effect of magnifying the scaffold. Meanwhile, the inner diameter and area of the scaffold cavity decrease, and the CT value of the scaffold increases 13 .…”
Section: Discussionmentioning
confidence: 99%