Design study of a PET detector with 0.5 mm crystal pitch for high-resolution preclinical imaging

Abstract: Preclinical positron emission tomography (PET) is a sensitive and quantitative molecule imaging modality widely used in characterizing the biological processes and diseases in small animals. The purpose of this study is to investigate the methods to optimize a PET detector for high-resolution preclinical imaging. The PET detector proposed in this study consists of a 28 × 28 array of LYSO crystals 0.5 × 0.5 × 6.25 mm3 in size, a wedged lightguide, and a 6 × 6 array of SiPMs 3 × 3 mm2 in size. The simulation res… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
7
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 10 publications
(7 citation statements)
references
References 32 publications
0
7
0
Order By: Relevance
“…All crystals could be clearly distinguished, as the size of the LYSO array was about 1 mm smaller than that of the SiPM array on each side, and a DOI resolution ∼2.0 mm was achieved. Zhang et al 40 optimized a high-resolution PET detector with a LYSO crystal of 0.5 × 0.5 × 6.25 mm 3 size and performed initial image studies with a prototype PET scanner consisting of 12 optimized detectors. However, for all the prototype small animal PET scanners developed so far with spatial resolution approaching 0.5 mm, the sensitivity was still much lower than current commercial small animal PET scanners.Lai et al 41 performed Monte Carlo simulations to design small animal PET scanners that can achieve ∼0.5 mm spatial resolution and high sensitivity with different scanner geometries and crystal lengths.…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…All crystals could be clearly distinguished, as the size of the LYSO array was about 1 mm smaller than that of the SiPM array on each side, and a DOI resolution ∼2.0 mm was achieved. Zhang et al 40 optimized a high-resolution PET detector with a LYSO crystal of 0.5 × 0.5 × 6.25 mm 3 size and performed initial image studies with a prototype PET scanner consisting of 12 optimized detectors. However, for all the prototype small animal PET scanners developed so far with spatial resolution approaching 0.5 mm, the sensitivity was still much lower than current commercial small animal PET scanners.Lai et al 41 performed Monte Carlo simulations to design small animal PET scanners that can achieve ∼0.5 mm spatial resolution and high sensitivity with different scanner geometries and crystal lengths.…”
Section: Introductionmentioning
confidence: 99%
“…All crystals could be clearly distinguished, as the size of the LYSO array was about 1 mm smaller than that of the SiPM array on each side, and a DOI resolution ∼2.0 mm was achieved. Zhang et al 40 . optimized a high‐resolution PET detector with a LYSO crystal of 0.5 × 0.5 × 6.25 mm 3 size and performed initial image studies with a prototype PET scanner consisting of 12 optimized detectors.…”
Section: Introductionmentioning
confidence: 99%
“…where S in,i is the energy signal read from the ith SiPM and N th is the preset threshold. Flood histograms calculated using threshold values ranging from 1% to 3.5% were assessed to determine the best threshold value [44].…”
Section: Discussionmentioning
confidence: 99%
“…When the threshold was set higher than 2%, the peak-to-valley ratio remained almost constant for both layers. Based on our previous experience [44], the optimal threshold was 2%, because higher thresholds cause distortion of the signal and decoding specks. The average peak-to-valley ratios at a threshold value of 2% for top and bottom crystals were 3.56 and 4.73, respectively.…”
Section: Flood Histograms Of Toray Reflectorsmentioning
confidence: 99%
“…27 In this work, theoretical calculations and experimental measurements were performed to compare the quality of the flood histograms generated by the dual-ended readout PET detectors obtained with the AM and EWM methods. Metrics such as contrast 28 and peak-to-valley ratio (PVR) 29,30 of the one-dimensional (1D) profiles across flood histograms and ratio of the width of the flood histograms of individual crystals to the distance between the locations of neighboring crystals in the flood histogram 31,32 were considered to compare the quality of the obtained flood histograms between the above two methods.…”
Section: Introductionmentioning