2007
DOI: 10.1088/0031-9155/52/5/015
|View full text |Cite
|
Sign up to set email alerts
|

Determination of parameters for a multiple-source model of megavoltage photon beams using optimization methods

Abstract: Accurate modelling of the radiation output of a medical linear accelerator is important for radiotherapy treatment planning. The major challenge is the adjustment of the model to a specific treatment unit. One approach is to use a multiple-source model containing a set of physical parameters. In this work, the parameters were derived from standard beam data measurements using optimization methods. The source model used includes sub-sources for bremsstrahlung radiation from the target, extra-focal photon radiat… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
69
0

Year Published

2014
2014
2019
2019

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 48 publications
(69 citation statements)
references
References 34 publications
0
69
0
Order By: Relevance
“…It consists of primary photon source, extrafocal photon source, electron contamination source, and photons scattered from wedges. Even though Acuros XB and AAA share the same multiple‐source model, the model parameters are different due to the differences in the dose calculation 11 , 36 …”
Section: Methodsmentioning
confidence: 99%
“…It consists of primary photon source, extrafocal photon source, electron contamination source, and photons scattered from wedges. Even though Acuros XB and AAA share the same multiple‐source model, the model parameters are different due to the differences in the dose calculation 11 , 36 …”
Section: Methodsmentioning
confidence: 99%
“…Details of the most important incremental improvements from the point of view of accurate dose calculation for small field sizes are presented in Torsti et al16 Currently, there are still limitations in the source model 19. It is important to remember that the head of the linac is not physically modeled.…”
Section: Methodsmentioning
confidence: 99%
“…The AcurosXB algorithm uses the multiple source model19 and was configured with the above mentioned set of input data. The modeling of the dynamic MLC deliveries is described through the use of two parameters related only to the energy, the leaf transmission (LT) and the dosimetric leaf gap (DLG) (which is the MLC gap assumed by Eclipse for closed MLC leaves to describe the transmission through the rounded MLC leaf tips).…”
Section: Methodsmentioning
confidence: 99%
“…The reader is encouraged to refer to the work of Tillikainen and colleagues (27) for an in‐depth explanation of this optimization process. For the primary photon source, a curve is fitted, which gives the mean energy as a function of radial distance from the CAX.…”
Section: Methodsmentioning
confidence: 99%