. Can. J. Chem. 61, 1949Chem. 61, (1983. A method has bcen developed for the routine radiochemical analysis of technetium-99 in human urinc. Thc excretcd technetium is coprecipitated from a 200 mL urine sample with CuS. Valcncc state adjustments are performed with K,SZOx. The precipitatc is dissolved in ammoniacal pcroxide and, after the Cu(ll) is removed by ion exchange, the Tc-99 is cxtractcd with 2-butanone (MEK). Subsequent evaporation of the organic solvent gives an csscntially wcightlcss source that if counted with a gas proportional counter will give a better limit of detection than liquid scintillation methods. The recovery of Tc-99 using this procedure is 80% ( u = 4%) and the limit of detection is 20 mBq (0.6 pCi) at the 95% confidence level for a routinc sample -counting time 20 minutes, counting cfficicncy 24%, ambient background 1.3 counts per minute (cpm).GARY H. KRAMER. Can. J. Chem. 61, 1949Chem. 61, (1983 On a mis au point une methode d'analyse radiochimique courante du teehnttium-99 dans I'urine humaine. A partir d'un Cchantillon de 200 mL d'urine, on fait coprtcipiter le technetium excrCtC avec du CuS. On rCalise des ajustements de la valence avec du K2S208 On dissout le prCcipit6 dans du peroxyde ammoniacal et on extrait le Tc-99 avec la butanone-2 (MEK) aprks avolr enlev6 le Cu(I1) par Ccl~ange d'ions. L'Cvaporation subsequente du solvant organique donne essentiellement une source imponderable qui si elle est comptCe avec un compteur proportionnelle de gaz donnera une meilleure lim~te de detection que les mCthodes de scintillation liquide. Par cette methode on rkcupkre le Tc-99 h 80% ( u = 4%) et la limite de detection est de 20 mBq (0,6 pCi) avec une certitude de 95% pour un eehantillon courant. LC temps de comptage est dc 20 minutes, I'efficacite de comptage est de 24%, le bruit de fond est de 1.3 compte par minute (cpm).[Traduit par le journal]