Development and evaluation of relative QA dosimeter for electron beam based on CsPbBr3
Seung-woo Yang,
Yo-han Shin,
Jae-hoon Jung
et al.
Abstract:Quality assurance (QA) of electron beams must be performed regularly to ensure accurate radiation therapy. Therefore, a QA dosimeter for electron beams is required for accurate electron beam QA measurement. In this study, the electron beam detection performance of CsPbBr3 material was measured. And it was evaluated whether it could be applied as a QA dosimeter for electron beams. To evaluate feasibility, reproducibility, linearity, and PDD were measured, and evaluated at 6 MeV, 9 MeV, and 12 MeV energies of th… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.