Abstract:We describe two test benches that were designed and constructed to perform a set of acceptance tests for about 1200 Photomultiplier Tube (PMT) units to be installed in the surface scintillation detectors of AugerPrime. Besides possessing robustness, long-term reliability, and low power consumption, each PMT-unit requires a dynamic range wide enough to cover signals ranging from a single to more than 20,000 minimum ionizing particles with not more than 5 % deviation from linear response. This posed a particular… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.