Abstract: The purpose of this study is to evaluate RadCalc, an independent dose verification software, for patient-specific quality assurance (PSQA) in online adaptive planning with a magnetic resonance linear accelerator (MR-linac) of a 1.5 T. Version 7.1.4 of RadCalc to introduce the capability to establish a beam model that incorporates MR field characteristics. A total of six models were established, with one using manufacturer-provided data and the others differing in percentage depth dose (PDD) data sources. Overa… Show more