2020
DOI: 10.35848/1347-4065/ab887e
|View full text |Cite
|
Sign up to set email alerts
|

Effect of methods for determining the frequency dependency of hydrophone sensitivity phase on instantaneous acoustic pressure of diagnostic ultrasound

Abstract: Ultrasonic diagnostic equipment must balance advanced performance and the safety of patients. Precise techniques have been investigated for determining instantaneous acoustic pressure of broadband ultrasound radiated from the equipment. These techniques require frequency dependency of the amplitude and phase in a hydrophone sensitivity. Phase response can be a constant value or determined by methods such as theoretical calculation and calibration. However, few studies have focused on quantitative evaluation of… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
8
0

Year Published

2021
2021
2025
2025

Publication Types

Select...
4

Relationship

2
2

Authors

Journals

citations
Cited by 4 publications
(8 citation statements)
references
References 36 publications
0
8
0
Order By: Relevance
“…The deconvolution method essentially includes the interpolation and extrapolation of the frequency response of hydrophone sensitivity and the frequency filtering of the instantaneous acoustic pressure waveform. 20) The frequency response of hydrophone sensitivity within the frequency range of certificated sensitivities was interpolated using a cubic spline, as used in previous studies, 10,17) to compensate for the mismatch between the frequency interval for the calibration frequency (1 MHz) and that for the frequency spectrum ( f s /N = 0.2 MHz).…”
Section: Signal Processing For Deconvolution Methodsmentioning
confidence: 99%
See 3 more Smart Citations
“…The deconvolution method essentially includes the interpolation and extrapolation of the frequency response of hydrophone sensitivity and the frequency filtering of the instantaneous acoustic pressure waveform. 20) The frequency response of hydrophone sensitivity within the frequency range of certificated sensitivities was interpolated using a cubic spline, as used in previous studies, 10,17) to compensate for the mismatch between the frequency interval for the calibration frequency (1 MHz) and that for the frequency spectrum ( f s /N = 0.2 MHz).…”
Section: Signal Processing For Deconvolution Methodsmentioning
confidence: 99%
“…This filter was used in previous studies. 10,17) The cutoff frequency for the two extrapolation methods was fixed at 40 MHz independent of f H , whereas that for the filtering treatment was f H , which varies according to the frequency range of the given certificated sensitivities.…”
Section: K F | ( )| and Phase K Fmentioning
confidence: 99%
See 2 more Smart Citations
“…The research group of the authors has been studying measurement techniques for ultrasonic sound field necessary to ensure the safety of medical ultrasonic equipment. [9][10][11][12][13][14][15][16][17][18][19][20] As for ultrasonic diagnostic equipment, a precise and practical measurement technique for instantaneous acoustic pressure of broadband ultrasound is required to balance the safety of patients in diagnosis 21,22) and the improvement of clarity of a diagnostic image. 19) Thus, a method using both amplitude and phase frequency responses of hydrophone sensitivity (hereinafter, called deconvolution method) [17][18][19][20][23][24][25][26][27][28] has been investigated.…”
Section: Introductionmentioning
confidence: 99%