2011
DOI: 10.1002/ccd.23028
|View full text |Cite
|
Sign up to set email alerts
|

Electrocardiographic changes and arrhythmias following percutaneous atrial septal defect and patent foramen ovale device closure

Abstract: We report a low risk of clinically significant post-procedure arrhythmias after device placement. Clinically significant heart block occurred in only two patients (0.3%). Changes in several markers of atrial conduction were found, suggesting an effect of device closure on intra-atrial conduction.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

3
42
0
2

Year Published

2013
2013
2018
2018

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 67 publications
(47 citation statements)
references
References 29 publications
3
42
0
2
Order By: Relevance
“…Among other findings, the average P-wave duration was increased on intermediate follow-up as compared to early follow-up (P < 0.001). Development of new-onset 1st degree AV Block after the procedure was associated with an increased risk of ATs post-procedure (P < 0.0001) [38].…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…Among other findings, the average P-wave duration was increased on intermediate follow-up as compared to early follow-up (P < 0.001). Development of new-onset 1st degree AV Block after the procedure was associated with an increased risk of ATs post-procedure (P < 0.0001) [38].…”
Section: Discussionmentioning
confidence: 99%
“…Johnson et al [38] reported an incidence of 5.2% of arrhythmias (32 patients among 610 patients) in the 4 months following device placement, including 29 patients with atrial tachyarrhythmias (22 atrial fibrillation, 7 atrial flutter), 1 with junctional tachycardia, and 2 with heart block. Among other findings, the average P-wave duration was increased on intermediate follow-up as compared to early follow-up (P < 0.001).…”
Section: Discussionmentioning
confidence: 99%
“…After the interventional treatment of ASDs, atrial arrhythmias are relatively common, including atrial tachycardia and premature atrial beating; the incidence rates for these arrhythmias range from 5.2 to 16.0% [14,15]. When there is a left-to-right shunt, preoperative right ventricular overload can occur, while a right ventricular volume load drop can occur postoperatively.…”
Section: Discussionmentioning
confidence: 99%
“…This does not often occur immediately, but will occur a month after the closure procedure as the myocardial edema subsides and the heart rate slows to facilitate recovery. The PR interval and the duration of the P wave was lengthened in the follow-up period, which indicated that the atrial conduction delay was prolonged, so it was necessary to extend the follow-up time, which was favorable for the discovery of potential atrioventricular block [14]. …”
Section: Discussionmentioning
confidence: 99%
“…İşlem esnasında ve sonrasında izlemlerde aritmi sıklığı %5 olarak bildirilmiştir. 17 Aritmi sıklığındaki artış interatriyal septumun cihazın santral bel kısmı tarafından gerilmesine bağlı olabileceği düşünül-mektedir. Bizim çalışmamızda da bir hastada geçici 2. derece atriyoventriküler blok gelişti.…”
Section: Perkütan Kapatma Işlemiunclassified