Electronics for Ultrasonic Imaging System

Abstract: Design of ultrasonic imaging system is presented. System has a modular structure with main acquisition and front end electronics separated in order to have minimal path for host PC connectivity and shortest path to ultrasonic transducer. Such acquisition modules placement allows reducing the induced EMI and increasing the flexibility of the system. Positioning module is also separate and allows various scanning equipment configurations. Evaluation of excitation and reception electronics parameters is presented… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
25
0

Year Published

2014
2014
2023
2023

Publication Types

Select...
4
2

Relationship

1
5

Authors

Journals

citations
Cited by 16 publications
(25 citation statements)
references
References 20 publications
0
25
0
Order By: Relevance
“…Attenuation was measured using toneburst signals and sine wave correlation (SWC) technique [12]. Sine and cosine function are fitted to the sampled signal s 1 …s M : were M is samples number, f s is sampling frequency and t 1 …t M are sampling time instances.…”
Section: B Attenuation Measurementmentioning
confidence: 99%
See 2 more Smart Citations
“…Attenuation was measured using toneburst signals and sine wave correlation (SWC) technique [12]. Sine and cosine function are fitted to the sampled signal s 1 …s M : were M is samples number, f s is sampling frequency and t 1 …t M are sampling time instances.…”
Section: B Attenuation Measurementmentioning
confidence: 99%
“…Excitation was performed using a unipolar pulser, capable of rectangular pulses production [12]. Two 2 MHz center frequency ultrasonic transducers TF2C6N from Dopler Inc. were used.…”
Section: A Experimental Setupmentioning
confidence: 99%
See 1 more Smart Citation
“…The polished 30 mm height and 20 mm diameter stainless steel cylinder was immersed in water tank (refer Fig. 1) and reflections collected using ultrasonic data acquisition system [8]. Transducer excitation was done using a unipolar pulser, using 10 V rectangular pulse of 50 ns duration for 10 MHz transducers and 100 ns duration for 5 MHz transducers.…”
Section: The Ultrasonic Signalsmentioning
confidence: 99%
“…If probing signal is of indefinite bandwidth such problem does not exist, but bandwidth is always limited in practice [6,7]. Furthermore, limited bandwidth is usually desirable, because signal acquisition system requirements can be relaxed [8], attenuation of ultrasound at high frequencies produces structural noise and signal degradation [9]. Deconvolution is used to address this problem [10][11][12].…”
Section: Introductionmentioning
confidence: 99%