Evaluating Factors Affecting Mean Glandular Dose in Mammography: Insights from a Retrospective Study in Dubai
Kaltham Abdulwahid Mohammad Noor,
Norhashimah Mohd Norsuddin,
Muhammad Khalis Abdul Karim
et al.
Abstract:Background/Objective: This study evaluates the mean glandular dose (MGD) in mammography screening for women aged 40–69 in Dubai, based on a retrospective analysis of a dose survey involving 2599 participants. Methods: MGD was calculated using the Dance formula. Results: The average MGD was 0.96 ± 0.39 mGy for mediolateral oblique (MLO) views and 0.81 ± 0.33 mGy for craniocaudal (CC) views. Weak inverse correlations were found between age and organ dose (OD) for both views, while a direct relationship was obser… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.