Abstract: Purpose: To explore candidate RayStation beam models to serve as a class-specific template for a TrueBeam treatment delivery system. Methods: Established validation techniques were used to evaluate three photon beam models: a clinically optimized model from the authors' institution, the built-in RayStation template, and a hybrid consisting of the RayStation template except substituting average MLC parameter values from a recent IROC survey. Comparisons were made for output factors, dose profiles from open fiel… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
7
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
4

Relationship

2
2

Authors

Journals

citations
Cited by 4 publications
(7 citation statements)
references
References 30 publications
0
7
0
Order By: Relevance
“…Once that was completed, points belonging to the tongue and groove zone were then assigned to one of the three subgroups: paired, exposed, or neglected. Similarly, specific points belonging to open field were reassigned to the calibration classification if their assignment changed after applying RayStation's position calibration 12 . The assignment criteria for the seven classifications are provided in Table 2.…”
Section: Methodsmentioning
confidence: 99%
See 3 more Smart Citations
“…Once that was completed, points belonging to the tongue and groove zone were then assigned to one of the three subgroups: paired, exposed, or neglected. Similarly, specific points belonging to open field were reassigned to the calibration classification if their assignment changed after applying RayStation's position calibration 12 . The assignment criteria for the seven classifications are provided in Table 2.…”
Section: Methodsmentioning
confidence: 99%
“…Similarly, specific points belonging to open field were reassigned to the calibration classification if their assignment changed after applying RayStation's position calibration. 12 The assignment criteria for the seven classifications are provided in Table 2.…”
Section: Scoringmentioning
confidence: 99%
See 2 more Smart Citations
“…In light of this, time may be better invested in creating a personalised, optimal model based on comprehensive commissioning of the machine. This has been shown as the best approach by Hansen et al [ 18 ] who compared a clinical model, a Raystation model and a hybrid IROC-Raystation model.…”
Section: Opening Statement – Jerome Gastaldomentioning
confidence: 99%