Abstract:In this study, we evaluated the stability and reliability of absorbed dose-to-water for an HDR 192 Ir sandwich setup phantom method by comparing measurements with absorbed dose-to-water determination based on the AAPM TG-43 protocol. Methods: The sandwich setup phantom was designed with a dedicated device for two ion chamber measurements of absorbed dose-to-water for a mHDR-v2r 192 Ir brachytherapy source is presented. To test the reliability of sandwich setup phantom of measurements with absorbed dose-to-wate… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.