Evaluation of the Reliability and Repeatability of Scheimpflug System Measurement in Keratoconus

Abstract: The level of repeatability of the measurements obtained with a Pentacam HR confirms the usefulness of this system for the diagnosis of keratoconus. Nevertheless, it has certain limitations for monitoring individual progression of keratoconus and planning intrastromal corneal ring segment insertion.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

1
22
2
1

Year Published

2019
2019
2021
2021

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 33 publications
(26 citation statements)
references
References 30 publications
1
22
2
1
Order By: Relevance
“…Many of the RLs found in the present study were lower than those in the de Luis Eguileor study 5 . This was especially true when all four of the devices image quality metrics were met.…”
Section: Discussioncontrasting
confidence: 86%
See 1 more Smart Citation
“…Many of the RLs found in the present study were lower than those in the de Luis Eguileor study 5 . This was especially true when all four of the devices image quality metrics were met.…”
Section: Discussioncontrasting
confidence: 86%
“…The Global Consensus on Keratoconus and Ectatic Disease defined ectatic progression by a consistent change, where the magnitude of the change is above the noise of the testing system 4 . In conforming to these guidelines, a study was published by de Luis Eguileor et al, which investigated the repeatability of measurements pertinent to the monitoring of progression in eyes with keratoconus using a Scheimpflug system (Pentacam HR) 5 . This study found that sequential measurements differing by more than 1.54 D, 1.08 D, and 0.71 ¼m in the steep keratometry, maximum keratometry, and vertical coma, respectively, likely represented real change.…”
Section: Introductionmentioning
confidence: 99%
“…The technical features of this system have been described elsewhere. 5 We used the default setting of 25 pictures/s. The Nidek ARK 560-A (Nidek Co Ltd, Tokyo, Japan) is a combined refractometer and keratometer.…”
Section: Instrumentsmentioning
confidence: 99%
“…Several studies have been carried out to investigate RC in measurements of the parameters used in the detection of progressive keratoconus. [4][5][6][7] However, RC is usually calculated based on measurements on the same day, rather than on different days. 8 The evaluation of the interday repeatability in subjects with keratoconus is thus of the utmost importance, as progression is diagnosed over time.…”
mentioning
confidence: 99%
“…Classicamente, inicia-se na puberdade e progride até a quarta década de vida, quando geralmente se estabiliza. (3) O diagnóstico é feito por exame clínico e avaliação tomográfica da córnea. Os sinais tomográficos da doença incluem: aumento da curvatura anterior da córnea, astigmatismo irregular e assimétrico, afinamento estromal e aumento da curvatura posterior.…”
Section: Introductionunclassified