Experimental small field 6MV output ratio analysis for various diode detector and accelerator combinations

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

4
41
0

Year Published

2014
2014
2020
2020

Publication Types

Select...
9

Relationship

0
9

Authors

Journals

citations
Cited by 53 publications
(45 citation statements)
references
References 24 publications
4
41
0
Order By: Relevance
“…Diode E value is 1.8% larger than recommended and is practically same as value measured by Mack et al [13]. Several authors reported over-response of shielded and unshielded silicon detectors in small photon fields [20,21,23]. It was proposed that formalism for reference dosimetry of small and nonstandard fields should be applied for correction of output factors in that case [19,24e26].…”
Section: Corrected Output Factorssupporting
confidence: 52%
See 1 more Smart Citation
“…Diode E value is 1.8% larger than recommended and is practically same as value measured by Mack et al [13]. Several authors reported over-response of shielded and unshielded silicon detectors in small photon fields [20,21,23]. It was proposed that formalism for reference dosimetry of small and nonstandard fields should be applied for correction of output factors in that case [19,24e26].…”
Section: Corrected Output Factorssupporting
confidence: 52%
“…A silicon detector will over-respond to low-energy scattered photons due to photoelectric effect in silicon sensitive volume [20]. To reduce this effect in the Diode P a disk-shaped silicon chip is surrounded by metal shielding.…”
Section: Detectorsmentioning
confidence: 99%
“…3,6,11,12 Silicon detectors have a very high resolution; typically, unshielded p-type silicon diodes with very small active volume are used to perform small-field dosimetric characterization and output factor (OF) measurements. 13 However, the mass density of the sensitive volume and surrounding medium strongly influences the small-field measurements with ionization microchambers and diodes. 9,14 Owing to their water equivalence and good spatial resolution, plastic scintillation detectors (PSDs) could play an important role in small-field dosimetry.…”
Section: Introductionmentioning
confidence: 99%
“…The output factor of a radiation field is defined as the ratio of the absorbed dose in water at a point located at the isocentre for the given field, relative to that at the same point in the reference field. In small fields, the measurement of the output factor is subject to large uncertainties [2].Silicon diode detectors are commonly used in small field dosimetry due to their relatively small sensitive volumes and their high radiosensitivity. However in the measurement of output factors, diodes have been reported to over-respond in small fields.…”
mentioning
confidence: 99%