Experimental study on malfunction of pacemakers due to exposure to different external magnetic fields

Abstract: There is a need for further investigations on pacemaker interference caused by different external magnetic fields to ensure safe working environment to workers with a pacemaker.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
16
0
2

Year Published

2012
2012
2023
2023

Publication Types

Select...
9

Relationship

1
8

Authors

Journals

citations
Cited by 22 publications
(18 citation statements)
references
References 29 publications
0
16
0
2
Order By: Relevance
“…Our method of measuring the induced voltage and comparison with the performance limits of ISO 14117 is a way of risk assessment which enables a general statement about potential EMI with CIEDs because all manufacturers are well-advised to test their devices according to the product standard to benefit from the presumption of conformity given by harmonized standards. In most other studies only a limited number of CIEDs were investigated which often limits their outcomes to the tested devices [ 7 , 27 , 28 , 29 ].…”
Section: Discussionmentioning
confidence: 99%
“…Our method of measuring the induced voltage and comparison with the performance limits of ISO 14117 is a way of risk assessment which enables a general statement about potential EMI with CIEDs because all manufacturers are well-advised to test their devices according to the product standard to benefit from the presumption of conformity given by harmonized standards. In most other studies only a limited number of CIEDs were investigated which often limits their outcomes to the tested devices [ 7 , 27 , 28 , 29 ].…”
Section: Discussionmentioning
confidence: 99%
“…W większości sytuacji nie doszło do zaburzenia funkcji urządzeń w warunkach działania pola elektromagnetycznego na poziomie poniżej międzynarodowych norm obowiązujących dla stanowisk pracy. W niektórych warunkach mogło jednak dojść do uszkodzenia kardiostymulatora nawet przy zachowaniu tych norm -wpływ działania pola zależał od jego przebiegu, częstotliwości i natężenia -przy czym bardziej podatne na uszkodzenie były kardiostymulatory jednobiegunowe niż bipolarne (34).…”
Section: Nowoczesne Techniki Terapeutyczne W Kardiologiiunclassified
“…The man was a volunteer patient in our study on pacemaker/ICD interference from magnetic fields 4 . After learning about his incident, we decided to replicate the situation.…”
Section: Case Reportmentioning
confidence: 99%
“…The man was a volunteer patient in our study on pacemaker/ICD interference from magnetic fields. 4 After learning about his incident, we decided to replicate the situation. We conducted the experiment under the medical surveillance of a doctor specialized in pacemaker treatments.…”
Section: Case Reportmentioning
confidence: 99%