2010
DOI: 10.5757/jkvs.2010.19.1.001
|View full text |Cite
|
Sign up to set email alerts
|

Extending the Pressure Limit for Turbomolecular Pump up to 133 Pa by using Conductance-Reducer and Measuring the Pressure Differences in Vacuum Chamber

Abstract: A dynamic flow system has been developed which can be used for vacuum gauge calibration by comparison method-a calibration method in which the reading of the gauge under calibration is compared to another calibrated vacuum gauge called the "secondary standard"-and other vacuum-related experiments. The chamber of the calibration system is pumped by a turbomolecular pump (TMP), backed by a scroll pump. As maximum acceptable pressure at the inlet of a TMP is 0.1 Pa, above which the TMP decelerates, the pumping sp… Show more