1989
DOI: 10.1007/bf01244696
|View full text |Cite
|
Sign up to set email alerts
|

Extraction inductively coupled plasma-optical emission spectrometry (ICP-OES) determination of traces of phosphorus in tungsten

Abstract: A method for the separation and preconcentration of traces of phosphorus from tungsten was developed. Solid phase extraction of the phosphovanadomolybdate complex performed on a micro-column was applied. Phosphorus was determined by optical emission spectroscopy (OES) with inductively coupled plasma (ICP) excitation. A limit of detection of 0.4/~g/g P with respect to the solid phase is obtained. By directly coupling the extraction/elution step to the ICP instrument a detection limit of 0.06 #g/g P in W was ach… Show more