Hp(3) vs TLD-100 for eye lens dosimetry in interventional radiology procedures: a preliminary study
Francesco Manna,
Gianleonardo De Nardellis,
Patrizio Antonio Carmosino
et al.
Abstract:Following the 2013/59/EURATOM Directive, the Italian 101/2020 Decree Law lowered the annual limit on the equivalent dose to eye lens from 150 to 20 mSv for exposed workers, in order to limit the risk of radiation-induced cataract. Such a drastic reduction makes it very important to perform accurate dose monitoring for medical staff that works with ionizing radiation as interventional radiologists and cardiologists. The standard for occupational eye lens dosimetry consists into Hp(3) calibrated dosimeters place… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.