2019
DOI: 10.1051/bioconf/20191502030
|View full text |Cite
|
Sign up to set email alerts
|

Interlaboratory study of ethanol usage as an internal standard in direct determination of volatile compounds in alcoholic products

Abstract: A collaborative interlaboratory study on the method of direct quantitation of volatile compounds in spirit drinks and alcoholic products was conducted. The discussed method applies ethanol, the major volatile component of an alcoholic product, as an internal standard. In this study 9 laboratories from 4 different countries were supplied with standard solutions for gas chromatographic measurements. Five aqueous ethanol 40% (v/v) standard solutions containing target compounds in concentrations ranging from 10 mg… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

0
7
0

Year Published

2020
2020
2023
2023

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 12 publications
(7 citation statements)
references
References 7 publications
0
7
0
Order By: Relevance
“…where C st i and C st IS are concentrations of the i-th analyzed compound and IS, correspondingly; A st i and A st IS are the detector responses for the i-th analyzed compound and IS, correspondingly. In the case of using ethanol as an IS compound, its concentration in each and every prepared standard solution equals 789300 mg/L AA [13][14][15].…”
Section: Split Ratiomentioning
confidence: 99%
“…where C st i and C st IS are concentrations of the i-th analyzed compound and IS, correspondingly; A st i and A st IS are the detector responses for the i-th analyzed compound and IS, correspondingly. In the case of using ethanol as an IS compound, its concentration in each and every prepared standard solution equals 789300 mg/L AA [13][14][15].…”
Section: Split Ratiomentioning
confidence: 99%
“…• the developed method was validated in one laboratory (single-laboratory study) (Charapitsa et al 2019b); • the developed method was investigated in international interlaboratory study (in 9 laboratories from 4 countries) (Charapitsa et al 2019a); • the influence of matrix effects to the results, obtained for developed method, was estimated for 13 ethanol-containing matrices (absence of matrix effect was proved) (Charapitsa et al 2021). • the perspectives of ethanol usage as a reference substance for the quantification of volatile compounds in alcoholic products by GC-MS-developed method were investigated (Korban et al 2019(Korban et al , 2021.…”
Section: Introductionmentioning
confidence: 99%
“…The data, obtained in abovementioned single-laboratory (Charapitsa et al 2019b) and interlaboratory (Charapitsa et al 2019a) studies (the metrological characteristics of the method, such as repeatability, intermediate precision, reproducibility, trueness, uncertainty), was applicable to control the acceptability and trueness (according to the ISO 5725-6: 1994) of results of analysis of only model water-ethanol solutions. The results of estimation of metrological characteristics (relative standard deviation (RSD) of repeatability-RSD r ; RSD of intermediate precision-RSD I(TO) ; relative expanded uncertainty P = 95%, with coverage factor k = 2-U; limit of detection-LOD; and limit of quantitation-LOQ) of the developed method, obtained in singlelaboratory study (Charapitsa et al 2019b) for concentration range from 2.10 to 5000 mg (L AA) −1 , are presented in Table 1.…”
Section: Introductionmentioning
confidence: 99%
“…When using FID, a large area of ethanol signal did not cause any inconveniences in the calculations. Although its peak area was more than 100 000 times greater than that of the analysed volatiles, ethanol was successfully employed in direct quantification . A specific procedure should be employed for the MS.…”
Section: Introductionmentioning
confidence: 99%
“…Although its peak area was more than 100 000 times greater than that of the analysed volatiles, ethanol was successfully employed in direct quantification. [9][10][11][12][13][14] A specific procedure should be employed for the MS. Here, we had to tune the ethanol peak area within the linear dynamic range of the MS detector.…”
mentioning