2017
DOI: 10.1093/jat/bkx062
|View full text |Cite
|
Sign up to set email alerts
|

LC–MS-MS with Post-Column Reagent Addition for the Determination of Zolpidem and its Metabolite Zolpidem Phenyl-4-carboxylic Acid in Oral Fluid after a Single Dose

Abstract: A rapid, selective and sensitive LC-MS-MS method with a post-column addition of acetonitrile was developed and fully validated for the quantitative determination of zolpidem and its major metabolite, zolpidem phenyl-4-carboxylic acid (ZPCA), in oral fluid. Preliminary sample treatment was limited to a simple dilution of 1 mL oral fluid specimen aliquots with methanol. Chromatography was performed on a Capcell Pak C18 MGII column (250 × 2.0 mm, 5 μm i.d., Shiseido, Tokyo, Japan) with isocratic elution using a w… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
7
0
1

Year Published

2019
2019
2023
2023

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 9 publications
(8 citation statements)
references
References 36 publications
0
7
0
1
Order By: Relevance
“…The following conditions were used: column, Shim-pack Scepter C18-120 (3 μm, 2.1 × 150 mm; Shimadzu), solvent A: 0.2% formic acid, solvent B: 0.2% (v/v) formic acid in 10% (v/v) acetonitrile, 40% (v/v) methanol, flow rate 0.25 mL/min, column temperature, 45°C. For stabilizing the ionization, 0.2 mL/min of acetonitrile was added post-column via a microvolume Tee-connector (Valco Instruments, Houston, TX, USA) by an additional LC-20AD pump [ 45 47 ]. Samples were measured in a randomized order.…”
Section: Methodsmentioning
confidence: 99%
“…The following conditions were used: column, Shim-pack Scepter C18-120 (3 μm, 2.1 × 150 mm; Shimadzu), solvent A: 0.2% formic acid, solvent B: 0.2% (v/v) formic acid in 10% (v/v) acetonitrile, 40% (v/v) methanol, flow rate 0.25 mL/min, column temperature, 45°C. For stabilizing the ionization, 0.2 mL/min of acetonitrile was added post-column via a microvolume Tee-connector (Valco Instruments, Houston, TX, USA) by an additional LC-20AD pump [ 45 47 ]. Samples were measured in a randomized order.…”
Section: Methodsmentioning
confidence: 99%
“…A post‐column addition of acetonitrile was achieved with a separate isocratic pump of the UltiMate ® 3000 Intelligent LC series (Thermo Scientific Dionex, Idstein, Germany). Instrument parameters were as described previously . Two calibration curves, using concentrations from 0.05 ng/mL to 200 ng/mL for zolpidem and from 0.1 ng/mL to 200 ng/mL for ZPCA, were employed to extend the assay's dynamic range.…”
Section: Methodsmentioning
confidence: 99%
“…Eligible volunteers received a single oral dose of zolpidem tartrate tablets (10 mg, Sanofi-Aventis Minsheng Pharmaceutical Co., Ltd, Hangzhou, China, Product lot: 9T034). Oral fluids were sampled before and at 1,2,3,4,5,6,7,8,9,10,11,12,13,14,15,16,24,30,36,42,48,54,60,66, and 72 hours after dosing.…”
Section: Characteristics Of Volunteers Were Detailed Inmentioning
confidence: 99%
See 2 more Smart Citations