Low-dose CT-based implant motion analysis is a precise tool for early migration measurements of hip cups: a clinical study of 24 patients

Abstract: Olivecrona (2020) Low-dose CT-based implant motion analysis is a precise tool for early migration measurements of hip cups: a clinical study of 24 patients,

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

2
35
0

Year Published

2020
2020
2024
2024

Publication Types

Select...
9

Relationship

4
5

Authors

Journals

citations
Cited by 24 publications
(37 citation statements)
references
References 14 publications
2
35
0
Order By: Relevance
“…We also included precision measurement of 9 cemented cups as these were visible in the CT stacks. That precision, 0.1-0.2 mm and 0.1-0.2°, is similar to the precision reported in another recently published article on the precision of CTMA for cups: 0.1-0.3 mm and 0.2-0.4° (Brodén et al 2020b).…”
Section: Discussionsupporting
confidence: 90%
“…We also included precision measurement of 9 cemented cups as these were visible in the CT stacks. That precision, 0.1-0.2 mm and 0.1-0.2°, is similar to the precision reported in another recently published article on the precision of CTMA for cups: 0.1-0.3 mm and 0.2-0.4° (Brodén et al 2020b).…”
Section: Discussionsupporting
confidence: 90%
“… 2016 ) and 0.2–2.3 mSv for a clinical hip study (Brodén et al. 2020 ). However, the range in radiation dose in this clinical study is quite wide.…”
Section: Discussionmentioning
confidence: 99%
“… 2016 , Brodén et al. 2019, 2020 ). However, to our knowledge there is no clinical study directly comparing CT and RSA in terms of precision for migration measurements on the same subjects.…”
mentioning
confidence: 98%