Measurement of Total Calcium in Serum by Atomic Absorption Spectrophotometry, with Use of a Strontium Internal Reference

Abstract: A method for total calcium in serum was optimized, a dual double-beam atomic absorption spectrophotometer being used, with strontium as an internal reference. The effect of HCl and lanthanum concentrations on the signals of both calcium and strontium was determined. The final analysis was unaffected by known sources of interference in serum at a 50-fold dilution of serum in 50 mmolar HCl containing 10 mmol LaCl3 and 0.12 mmol of SrCl2 per liter. Within-run precision for 30 replicate analyses performed during 3… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
15
0
1

Year Published

1976
1976
2024
2024

Publication Types

Select...
5
2
1

Relationship

0
8

Authors

Journals

citations
Cited by 86 publications
(16 citation statements)
references
References 0 publications
0
15
0
1
Order By: Relevance
“…Aliquots were taken for the measurement of both calcium concentration and Ta-radioactivity. Calcium concentration was analyzed by an atomic absorbtion spectrophotometer (Unicam SP90A) applying the method of Pybus et al (1970). '"Ca-activity was determined by the liquid scintillation counter by setting the discriminators to the energy level where no "-activity was present.…”
Section: Animals and Exprrimenrcil Conditionsmentioning
confidence: 99%
“…Aliquots were taken for the measurement of both calcium concentration and Ta-radioactivity. Calcium concentration was analyzed by an atomic absorbtion spectrophotometer (Unicam SP90A) applying the method of Pybus et al (1970). '"Ca-activity was determined by the liquid scintillation counter by setting the discriminators to the energy level where no "-activity was present.…”
Section: Animals and Exprrimenrcil Conditionsmentioning
confidence: 99%
“…method for total calcium and sodium in serum [3,4]. For calcium, the analytical procedure was based on the flame atomic absorption spectrometr ic (FAAS) method of Pybus, Feldman, and Bowers [5] , while for sodium it was based on…”
Section: Statistical Correlations Arementioning
confidence: 99%
“…(3) wipe off the tip with a clean, absorbent paper, (4) tilt the pipet to a horizontal position, (5) allow a small volume of air to leak in and rotate the pipet so that the conditioning liquid wets all the internal surface to approximately 0.5 cm above the calibration mark, (6) discard this conditioning solution, and 7repeat steps (1)(2)(3)(4)(5)(6) .…”
Section: Bmentioning
confidence: 99%
See 1 more Smart Citation
“…Interferências químicas provocadas pela matriz, devido à presença de elementos facilmente ionizáveis49 . 5.…”
unclassified