2018
DOI: 10.17073/1683-4518-2018-12-44-48
|View full text |Cite
|
Sign up to set email alerts
|

Metakaolin test for the chemically bound water

Abstract: The technique is presented for the chemically bound water determination in metakaolin (the mass fraction of water being within 0,5 and 5 %) using the ELTRA (Germany) CW-800M infrared spectrometer analyzer at 900 °C. The analysis of one sample lasts within 5 and 6 minutes, this allows to control for the metakaolin the incoming samples and not to pass the defective product into operation. It was shown that the analysis for the water can be carried out at 600°C, given that the analysis lasts 20‒22 minutes. The de… Show more