Micro-Milli Scale for the Removal of Iron and Copper by a Positively
Charged Foam before Nickel Colorimetric Detection using Handy
Spectrometer
Piyanat Issarangkura Na Ayutthaya,
Monnapat Vongboot
Abstract:Introduction:
Simple alternative detection of nickel with iron and copper as interferences was proposed. The procedure was based on sampling a milligram sample and a micro-milliliter operation.
Methods:
The method has been applied to jewelry items. A 50 mg sample was digested by 5 mL of nitric acid with heating. The obtained solution was added with KSCN before passing through polyurethane foam (PUF) (1 cm i.d. × 8 cm length). Some metal ions-SCN complexed (e.g., Fe(III) and Cu(II)) were retained in the colum… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.