2024
DOI: 10.3389/fnins.2024.1324971
|View full text |Cite
|
Sign up to set email alerts
|

Objective preclinical measures for bone conduction implants

Irina Wils,
Alexander Geerardyn,
Tristan Putzeys
et al.

Abstract: The study evaluates the accuracy of predicting intracochlear pressure during bone conduction stimulation using promontory velocity and ear canal pressure, as less invasive alternatives to intracochlear pressure. Stimulating with a percutaneous bone conduction device implanted in six human cadaveric ears, measurements were taken across various intensities, frequencies, and stimulation positions. Results indicate that intracochlear pressure linearly correlates with ear canal pressure (R2 = 0.43, RMSE = 6.85 dB),… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...

Citation Types

0
0
0

Year Published

2024
2024
2025
2025

Publication Types

Select...
3

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
references
References 38 publications
0
0
0
Order By: Relevance