On the Use of the Fricke-Pluronic F-127 Gel Dosimeter for Radiation Isocenter Testing of a Medical Linear Accelerator

Michał Piotrowski,
Piotr Maras,
Marek Kozicki

Abstract: This work presents a Fricke-XO-Pluronic F-127 2D radiochromic dosimeter with a flat-bed scanner for 2D reading and a dedicated data processing software package as a tool for performing coincidence testing of the radiation and mechanical isocenter of a medical accelerator. The optimal irradiation parameters were determined as follows: monitor units per beam and multi-leaf collimator gap, which are ≤750–≤2500 MU and 2–5 mm, respectively, for a cuboidal container with dimensions of 12 × 12 × 0.3 cm3. Despite the … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
6
0

Year Published

2024
2024
2024
2024

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
(6 citation statements)
references
References 39 publications
0
6
0
Order By: Relevance
“…Also, the irradiated dark blue zones are unstable and disperse over time after irradiation. This can be seen with the naked eye just 2.5 h after irradiation and storage at room temperature (see Figure 2 in [ 20 ]). Therefore, the Fricke–XO–Pluronic F-127 cannot be used as a reusable dosimeter.…”
Section: Discussion: Nbt–pluronic F-127 Vs Fricke–xo–pluronic F-127mentioning
confidence: 98%
See 4 more Smart Citations
“…Also, the irradiated dark blue zones are unstable and disperse over time after irradiation. This can be seen with the naked eye just 2.5 h after irradiation and storage at room temperature (see Figure 2 in [ 20 ]). Therefore, the Fricke–XO–Pluronic F-127 cannot be used as a reusable dosimeter.…”
Section: Discussion: Nbt–pluronic F-127 Vs Fricke–xo–pluronic F-127mentioning
confidence: 98%
“…To date, we have investigated two radiochromic gel dosimeters as 2D systems for routine medical linear accelerator (X-ray radiation) testing–coincidence of mechanical and radiation isocenters: NBT–Pluronic F-127 in current work and Fricke–XO–Pluronic F-127 reported elsewhere [ 20 ]. There is only one similarity between these dosimeters: the Pluronic F-127 matrix.…”
Section: Discussion: Nbt–pluronic F-127 Vs Fricke–xo–pluronic F-127mentioning
confidence: 99%
See 3 more Smart Citations