Performance and safety of the PRICO closed-loop oxygen saturation targeting system in neonates: pragmatic multicentre cross-over study (TarOx Study)
Maria Wilinska,
Thomas Bachman,
Tomasz Szczapa
et al.
Abstract:ObjectiveThis study aims to evaluate the performance of the fabian-Predictive-Intelligent-Control-of-Oxygenation (PRICO) system for automated control of the fraction of inspired oxygen (FiO2).DesignMulticentre randomised cross-over study.SettingFive neonatal intensive care units experienced with automated control of FiO2and the fabian ventilator.Patients39 infants: median gestational age of 27 weeks (IQR: 26–30), postnatal age 7 days (IQR: 2–17), weight 1120 g (IQR: 915–1588), FiO20.32 (IQR: 0.22–0.43) receivi… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.