Purpose: To investigate the correlation between the preoperative biometric parameters of the anterior segment and the vault after implantable Collamer lens (ICL) implantation via this retrospective study. Design: Retrospective clinical study. Methods: A total of 78 eyes from 41 patients who underwent ICL implantation surgery were included in this study. Preoperative biometric parameters, including white-to-white (WTW) diameter, central corneal thickness, keratometer, pupil diameter, anterior chamber depth, sulcus-to-sulcus diameter, anterior chamber area (ACA) and central curvature radius of the anterior surface of the lens (Lenscur), were measured. Lenscur and ACA were measured with Rhinoceros 5.0 software on the image scanned with ultrasound biomicroscopy (UBM). The vault was assessed by UBM 3 months after surgery. Multiple stepwise regression analysis was employed to identify the variables that were correlated with the vault. Results: The results showed that the vault was correlated with 3 variables: ACA (22.4 ± 4.25 mm2), WTW (11.36 ± 0.29 mm) and Lenscur (9.15 ± 1.21 mm). The regressive equation was: vault (mm) = 1.785 + 0.017 × ACA + 0.051 × Lenscur - 0.203 × WTW. Conclusions: Biometric parameters of the anterior segment (ACA, WTW and Lenscur) can predict the vault after ICL implantation using a new regression equation.