2010
DOI: 10.1016/j.ijrobp.2009.10.040
|View full text |Cite
|
Sign up to set email alerts
|

Potentials and Limitations of Guiding Liver Stereotactic Body Radiation Therapy Set-Up on Liver-Implanted Fiducial Markers

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
70
0

Year Published

2011
2011
2024
2024

Publication Types

Select...
7
2

Relationship

1
8

Authors

Journals

citations
Cited by 82 publications
(71 citation statements)
references
References 16 publications
1
70
0
Order By: Relevance
“…Many studies on liver motion characterization using markers have been published. 12,16,19 Seppenwoolde et al showed that the tumor position can be accurately predicted using implanted markers considering that there is a real influence of the marker-tumor distance on the set-up accuracy. They also showed that the diaphragm dome was the second best predictor of tumor motion and that it could be used as a surrogate for tumor located close to it.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…Many studies on liver motion characterization using markers have been published. 12,16,19 Seppenwoolde et al showed that the tumor position can be accurately predicted using implanted markers considering that there is a real influence of the marker-tumor distance on the set-up accuracy. They also showed that the diaphragm dome was the second best predictor of tumor motion and that it could be used as a surrogate for tumor located close to it.…”
Section: Discussionmentioning
confidence: 99%
“…These can be used during the patient's setup and for the tumor motion verification as they are implanted inside or close to the target. 16 Recent report evaluated the geometric accuracy of the surrogate-based gated VMAT with a respiratory phantom and also on real patients cases including liver tumors. 17 Li et al showed in a phantom study a high geometric accuracy (average error of 0.8 mm in the CC direction) when no targetsurrogate relation changes occurred during the treatment.…”
Section: Introductionmentioning
confidence: 99%
“…Details about this procedure have been reported earlier. 5,[11][12][13] From 2011, only one large contrast enhanced planning CT in the venous phase was acquired. The tumor delineations have always been reviewed by an experienced radiologist.…”
Section: Treatment Preparation and Deliverymentioning
confidence: 99%
“…fiducial markers are implanted in proximity of the target lesion and can be visualized by computed tomography or x-ray images as a tumor surrogate to assure its correct position during treatment. New technologies such as gating and tracking of target lesions depend on the breathing phase and are correlated with the dose application [43][44][45].…”
Section: Discussionmentioning
confidence: 99%