2023
DOI: 10.1002/acm2.14094
|View full text |Cite
|
Sign up to set email alerts
|

Practical experience commissioning MRI‐compatible tandem and ring applicators for use with the Bravos HDR afterloader

Abstract: Five complete MR‐conditionally approved ring sets, including fifteen tandems, and two additional rings, were commissioned at an institution intending to use them in an MRI planning environment with a Bravos HDR brachytherapy remote afterloader. Channel length, radiograph, autoradiograph, ring offset, and treatment interrupt measurements were performed, and applicators were assessed in both CT and MRI. During commissioning, one ring was found to be defective and was returned to the manufacturer for a replacemen… Show more