Abstract:Conclusion:Comparison of synchronized 24-hour BP measurements on both arms showed large measurement uncertainties with possible implications on clinical grading of hypertension and dipping pattern. Exclusion of disturbed cuff pressure curves, not detected by the conventional software of BP devices, mitigates the effect but still results in worrying discrepancies. These results are an imperative
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.