Quantification and Dosimetric Impact of Normal Organ Motion During Adaptive Radiation Therapy Plan Verification Using a 1.5 Tesla Magnetic Resonance Equipped Linear Accelerator (MR LINAC)

Abstract: Results: A total of 2278 SE and DE image frames were analyzed. DE imaging resulted in significantly improved tracking vs. SE on 884/2278 (38.8%) of image frames. In those frames, the average tracking error was 1.78 +/-2.02 mm vs. 2.21 +/-2.38 mm, for DE vs. SE imaging, respectively (P < 0.001). For 4/10 patients, the average reduction in tracking error with DE vs. SE tracking was > 0.7 mm. No correlation was observed between tumor volume/location and improved tracking accuracy with DE imaging. Additionally, th… Show more