Quantitative and Compositional Analysis of Trace Amino Acids in Icy Moon Analogues Using a Microcapillary Electrophoresis Laser-Induced Fluorescence Detection System

Abstract: Enceladus’s jet activity, Europa’s putative plume activity, and the evidence supporting global subsurface oceans have made them high-priority targets for future NASA missions. In situ analysis of organic molecules in the jets/plumes or subsurface oceans of these icy moons would provide relevant, detailed information on the formation, habitability, and ongoing planetary processes of celestial bodies and could provide the first evidence of the potential for extant life beyond Earth. Microcapillary electrophoresi… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
7
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
4

Relationship

1
3

Authors

Journals

citations
Cited by 4 publications
(7 citation statements)
references
References 108 publications
0
7
0
Order By: Relevance
“…This method is also inherently less prone to electrodispersive effects, as it uses hydrodynamic injection as opposed to electrokinetic injection (see Supporting Information). Although previous CE-LIF analyses have reported a significant deterioration in signal and peak resolution at <5 mM Mg 2+ , we demonstrate that samples with Mg 2+ concentrations of >400 mM can be successfully analyzed for amino acid content using a simple running buffer consisting of only sodium tetraborate.…”
Section: Resultsmentioning
confidence: 54%
See 2 more Smart Citations
“…This method is also inherently less prone to electrodispersive effects, as it uses hydrodynamic injection as opposed to electrokinetic injection (see Supporting Information). Although previous CE-LIF analyses have reported a significant deterioration in signal and peak resolution at <5 mM Mg 2+ , we demonstrate that samples with Mg 2+ concentrations of >400 mM can be successfully analyzed for amino acid content using a simple running buffer consisting of only sodium tetraborate.…”
Section: Resultsmentioning
confidence: 54%
“…Capillary electrophoresis (CE) is a powerful separation technique that offers exceptionally high separation efficiencies, extremely small sample volumes and reagent consumption, and can be coupled to numerous detection systems . When coupled with a laser-induced fluorescence (LIF) detection system, CE can provide sub-nM (or mere attomoles of molecules) limits of detection (LODs). Multiple amine-reactive fluorophores have been developed, enabling quantitative compositional analysis of ultralow abundance amino acids via CE-LIF . While sample salinity can lead to electro-dispersive effects that reduce resolution and signal, optimization of methodological parameters has been shown to be effective at minimizing the deleterious effects of high salinity. , Multivalent cations are most problematic, as these can alter intercapillary surface chemistries, leading to electroosmotic flow (EOF) inhibition in addition to dispersive effects, resulting in drastically reduced peak resolution, sensitivity, and separation efficiency at cation concentrations as low as 5 mM .…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…To avoid or reduce the need for separate desalting steps, a high concentration (30-80 mM) of borate buffer is typically used and further improved by adding EDTA (Stockton et al, 2009a;Creamer et al, 2017). Duca et al (2022) used a 35 mM borate buffer as the background electrolyte, which allowed for a maximum of 1 mM H 2 SO 4 , 10 mM Na 2 CO 2 , and 10 mM MgSO 4 (with 5 mM EDTA) in the injected sample before separation failed when analyzing amino acids. Alternatively, high concentrations of acetic acid (1-5 M) can be used to enable analysis of samples with high ionic strength (Santos et al, 2018;Mora et al, 2022).…”
Section: Introductionmentioning
confidence: 99%
“…Alternatively, an injector valve can improve repeatability (Zamuruyev et al, 2021). ME, on the other hand, struggles with injection bias that consequently affects the measurement trueness (Duca et al, 2022). Furthermore, the channels often consist of two or more materials that may or may not be fully characterized, ultimately leading to unknown EOF characteristics.…”
Section: Introductionmentioning
confidence: 99%