Background
Advanced ablation strategies are needed to treat ventricular tachycardia (VT) and premature ventricular contractions (PVC) refractory to standard unipolar radiofrequency ablation (Uni-RFA). Bipolar radiofrequency catheter ablation (Bi-RFA) has emerged as a treatment option for refractory VT, PVC. Multicenter registry data on the use of Bi-RFA in the setting of refractory VT and PVC are lacking.
Purpose
The aim of this Bi-RFA registry is to determine its real-world safety, feasibility, and efficacy in patients with refractory VT/PVC.
Methods
Consecutive patients undergoing Bi-RFA at sixteen European centers for recurring VT/PVC after at least one standard Uni-RFA were included. Second ablation catheter was used instead of a dispersive patch and was positioned at the opposite site of the ablation target.
Results
Between March 2021 and August 2024, ninety-one patients underwent 94 Bi-RFA procedures (74 males, age 62±13, prior Uni-RFA range 1-8). Indications were recurrence of PVC (n=56), VT (n=20), electrical storm (n=13), or PVC-triggered ventricular fibrillation (n=2). Procedural time was 160±73min, Bi-RFA time 426±286s, mean Uni-RFA time 819±697s. Elimination of clinical VT/PVC was achieved in 67 (74%) patients, suppression of VT/PVC in a further 10 (11%) patients. In the remaining 14 patients (15%) no effect on VT/PVC was observed. Three major complications occurred: coronary artery occlusion, AV block and arteriovenous fistula. Follow-up lasted 7±8 months. Nineteen (61%) remained VT-free. ≥80% PVC burden reduction was achieved in 45 (78%).
Conclusions
This real-world registry data indicates that Bi-RFA appears safe, is feasible, and effective in the majority of patients with VT/PVC.