Abstract:Objectives: To compare the image quality, accuracy and effective dose of the default diagnostic program with the low dose program. Methods: The diagnostic and low dose programs of three-dimensional rotational angiography were compared considering their signal to noise ratio, their carrier to noise ratio and their actual radiation dose in μGy·m²/kg. An anthropomorphic phantom underwent both types of scans to evaluate the effective dose. Comparative measurements of vessel diameters were taken in the diagnostic a… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.