2021
DOI: 10.1038/s41598-021-81071-4
|View full text |Cite
|
Sign up to set email alerts
|

Simultaneous mapping of EMCD signals and crystal orientations in a transmission electron microscope

Abstract: When magnetic properties are analysed in a transmission electron microscope using the technique of electron magnetic circular dichroism (EMCD), one of the critical parameters is the sample orientation. Since small orientation changes can have a strong impact on the measurement of the EMCD signal and such measurements need two separate measurements of conjugate EELS spectra, it is experimentally non-trivial to measure the EMCD signal as a function of sample orientation. Here, we have developed a methodology to … Show more

Help me understand this report