Single-channel EEG signal extraction based on DWT, CEEMDAN, and ICA method

Abstract: In special application scenarios, such as portable anesthesia depth monitoring, portable emotional state recognition and portable sleep monitoring, electroencephalogram (EEG) signal acquisition equipment is required to be convenient and easy to use. It is difficult to remove electrooculogram (EOG) artifacts when the number of EEG acquisition channels is small, especially when the number of observed signals is less than that of the source signals, and the overcomplete problem will arise. The independent compone… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
2
0

Year Published

2023
2023
2024
2024

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(2 citation statements)
references
References 27 publications
0
2
0
Order By: Relevance
“…The independent component corresponding to the sample entropy values satisfying the Gomez Herrero condition was regarded as an artifact and set to zero ( Gomez Herrero et al, 2006 ). Finally, the inverse independent component analysis was then performed to reconstruct the new approximate and detail components, based on which the EEG signal with the rejection of cardiac field artifact, eye movement artifact, and muscle artifacts was obtained ( Hu et al, 2022 ) ( Supplementary Figures S1–S6 ) R-peak and T-peak were detected and labeled automatically with the help of peak detector functions integrated into Acqknowledge and Kubios HRV (Kubios 3.4, University of Eastern Finland, Kuopio) software. The interbeat interval (IBI) between 300 and 2000 ms, consecutive IBI differences ≤200 ms, and prolongations or shortenings ≤20% than the average of five preceding sinus rhythm IBIs were considered as sinus rhythm QRS complexes ( Liu et al, 2020 ).…”
Section: Methodsmentioning
confidence: 99%
“…The independent component corresponding to the sample entropy values satisfying the Gomez Herrero condition was regarded as an artifact and set to zero ( Gomez Herrero et al, 2006 ). Finally, the inverse independent component analysis was then performed to reconstruct the new approximate and detail components, based on which the EEG signal with the rejection of cardiac field artifact, eye movement artifact, and muscle artifacts was obtained ( Hu et al, 2022 ) ( Supplementary Figures S1–S6 ) R-peak and T-peak were detected and labeled automatically with the help of peak detector functions integrated into Acqknowledge and Kubios HRV (Kubios 3.4, University of Eastern Finland, Kuopio) software. The interbeat interval (IBI) between 300 and 2000 ms, consecutive IBI differences ≤200 ms, and prolongations or shortenings ≤20% than the average of five preceding sinus rhythm IBIs were considered as sinus rhythm QRS complexes ( Liu et al, 2020 ).…”
Section: Methodsmentioning
confidence: 99%
“…Djemal et al [27] combined ICA with the Regenerative Multi-Dimensional Singular Value Decomposition method. Hu et al [28] proposed a method that combines the discrete wavelet transform, complete empirical mode decomposition for adaptive noise, and ICA, which can effectively identify and remove EEG noise such as sharp signals and impulse interference. Liu et al [29] proposed a novel recurrent neural networkbased EEG state estimation model, which automatically replaces the detected noises with the estimated values to achieve a robust EEG perception.…”
Section: Introductionmentioning
confidence: 99%